Application Guide

Membrane & Filter Characterization

How capillary flow porometry, bubble point testing, and complementary methods characterize the through-pores of membranes and filters — and how to report the results consistently.

Why through-pores matter in filtration

A filter retains particles because of the narrowest constriction along each continuous path from one face to the other. Dead-end and closed pores contribute to total porosity but not to filtration, which is why membrane QC relies on porometry — methods that only see through-pores — rather than mercury intrusion.

The key quantities are the bubble point (the pressure at which gas first displaces the wetting liquid from the largest through-pore), the mean flow pore size (the pressure at which flow through the wet sample equals half the flow through the dry sample), and the through-pore size distribution between them.

Converting pressure to pore size

For a wetting liquid in a cylindrical pore, the Young–Laplace relation gives the diameter that empties at differential pressure ΔP:

d = 4 γ cos θ / ΔP  — with complete wetting (θ = 0°), d = 4 γ / ΔP. Example: Galwick (γ = 15.9 mN/m) with a bubble point of 100 kPa gives d = 4 × 0.0159 / 100,000 m ≈ 0.64 µm.

ASTM F316 instead writes d = Cγ/p with an empirical constant (C = 2860 for p in Pa, or 0.415 for p in psi, d in µm, γ in mN/m), which gives diameters about 0.715 times the ideal-cylinder value. Always state which convention and which wetting liquid were used; the two are not interchangeable.

Wetting liquidSurface tension (mN/m, approx.)Typical use
Galwick / Porewick (perfluorinated)15.9 / 16Default for most membranes; wets almost everything
Silwick (silicone oil)20.1Alternative low-surface-tension liquid
Isopropanol~21.7Common for integrity testing of hydrophobic filters
Water72.8 (20 °C)Hydrophilic sterilizing-grade filters; higher pressures for the same pore size

Choosing the right test

Membrane typeTypical methodNotes
Microfiltration (~0.1–10 µm)Gas–liquid capillary flow porometryBubble point, mean flow pore, distribution
Ultrafiltration (~2–100 nm)Liquid–liquid displacement porometry; MWCO by solute rejection (dextrans, PEG)Gas–liquid CFP needs very high pressures in this range
Sterilizing-grade filtersBubble point, diffusion or pressure-hold integrity testsCorrelated with bacterial retention (ASTM F838)
Nonwovens, geotextiles, sintered metalsCapillary flow porometry / bubble testASTM D6767 (geotextiles), ISO 4003 (sintered metals)

Standards

Questions about your material?

Use the calculators to convert test pressures to pore sizes, or email a short description of your sample and what you need to measure.